Our filtration devices have been rigorously tested and optimized to reduce process times, increase analytical efficiency, and produce consistent, measurable results. We offer polymeric, glass fiber, and ceramic membranes--in all standard diameters and a range of pore sizes--as disc/sheet filters or housed within syringe and capsule devices for single-use applications. Sterile and non-sterile options are available, along with a multitude of other custom features to accommodate end-users' diverse needs.

Frequently Asked Questions

Q. Will Sterlitech Track Etch filter membranes keep liquid behind the filter and let gases pass through?

A. PVP-Free Polycarbonate membranes have a water contact angle of approximately 90° and will not spontaneously wet out with liquids that have a surface tension equivalent to or greater than water (1 dyne). Due to the low water contact angle, polycarbonate membranes do not make effective vent filters. Low differential pressures will allow liquid water to break through the pores. We recommend membranes with a higher water entry pressure such as Hydrophobic PTFE, Hydrophobic Polyethylene, and Oleophobic Polyester for venting applications. Effective vent filters will allow permeation of gasses, while blocking liquid from entering the pores. Water vapor and other gases will pass through a hydrophobic vent membrane.

Q. Can Polycarbonate membrane be bonded with adhesives?

A. Polycarbonate membranes can be adhesive bonded to each other or to other plastics, metals, glass or wood using commercially available one component, two component and pressure adhesives.

Q. I am looking for a black polycarbonate membrane pore size that is not listed. Can I dye the standard polycarbonate membrane? How do I do this?

A. Yes, but we usually recommend the PVP-free polycarbonate membranes. Note that black polycarbonate membranes are used for counting particulates and performing fluorescent microscopy.

If you need to create a black membrane that is not listed on this site, the standard method for rendering polycarbonate membranes black in a lab follows:

Dissolve 2 gm of Irgalan black (Chemical Index, acid black 107) in 1 liter of 2% acetic acid.
Soak membranes in Irgalan black solution for 24 hours.
Rinse with water, air dry.
To obtain a darker black, dry at 180ºF (82ºC) for 15 minutes.

**Note: There have been instances of Irgalan Black "bleeding" off of the membrane.   This is not common, but has occurred.

Q. Why offer both polyester and polycarbonate track-etch filter membranes?

A. The greatest advantage of using polyester is that it has better solvent resistance and it is prone to fewer wrinkles in the manufacturing process. One of the main reasons polycarbonate was originally used and continues to be used by most people is that for over 30 years the end-users have known mainly polycarbonate.

Polycarbonate works better with live or fixed cells than does polyester. However, both polycarbonate  and polyester offer very high quality end product that is suitable for a large variety of applications.

Q. I am having a problem when I force liquids under pressure through a membrane with a very small pore size. Little or no liquid is seen to pass through the membrane, what could be the problem?

A. There could be a number of things going on to restrict the flow through the membrane.

First you want to take the original flow rate into consideration, for example a 25mm polycarbonate membrane with a 0.05 micron pore size and a flow rate of 0.4ml/min/cm2 is only going to get about 1.2 mil of liquid through a minute. This amount decreases with smaller pore sizes.

Another thing to consider is if the liquid is being properly prefiltered. With such small pore sizes, 0.05-0.01micron, you need to filter in a step down manner so as not to plug the pores. Step down means a series of filters with decreasing pore size.

This should eliminate most problems with flow through the membrane; other issues need to be evaluated on an individual basis.

Q. I just received a box of Sterlitech Track-Etch filter membranes and I am trying to use them, but there is no liquid passing through. What is wrong?

A. Believe it or not, in most cases end-users tend to use the blue separator paper instead of the membrane because the separator paper has the physical qualities that make it look more like a typical depth filter than the Track-Etch filter membrane.

Always remember that the filter membrane is usually opaque or translucent and is generally white or white with a yellow or green tint to it. The separator paper is usually blue or imprinted and should not be used for filtration.

Another reason for lack of filtration is due to particle loading on the membrane that would cause the filter to plug. Because Sterlitech Track-Etch filter membranes are made of plastic with cylindrical pores, they capture 100% of all particles larger than the pore size. Therefore, they have a higher incidence of plugging due to their high level of capturability. It is highly recommended that unless the end-user is capturing particles on the surface for analysis that a glass fiber prefilter is used on top of the track-etch membrane. The glass fiber prefilter will capture most of the larger particulate, thereby keeping it from loading up on the surface of the membrane and plugging it.

The use of a mesh spacer (drain disks) is also helpful for maximizing flow. A mesh spacer is set under the membrane and on top of the filter holder. The mesh spacer lifts the membrane off the mostly solid filter holder surface, allowing increased filtration by allowing tangential flow to occur. This is particularly applicable with stainless steel supports.

Q. We would like a higher density Polycarbonate membrane (PCTE). Is it available?

A. Yes, custom density PCTE is available.   Please contact our sales team for a quote ([email protected])

Q. What are the advantatges of Sterlitech polycarbonate and polyester membrane filters?

A. Yes, the polycarbonate track-etch (PCTE) membrane filters have good biocompatibility and pass USP Class VI testing. The PCTE membrane filters are neither cytotoxic nor bactericidal. Cells and bacteria will grow on the filters when proper nutrients are supplied and proper conditions are maintained.

Q. What membrane works best for cell studies?

A. Sterlitech Polycarbonate (PCTE) and Polyester (PETE) membranes are ideal for cell studies. They are neither cytotoxic nor bactericidal. Cells will grow on the membranes if provided with a nutrient.

The membranes can be repeatedly autoclaved at 121°C (250°F), and no damage has been observed in membranes with a sustained exposure to temperatures of 140°C (284°F) in air or steam.

PCTE membranes are neutral biologically, being neither cytotoxic nor bactericidal. Cells and bacteria will grow on PCTE membranes when proper nutrients are supplied. PCTE membranes pass all USP Class VI tests for bio-compatibility and are completely safe to use in implant studies.

Randomly selected samples of polycarbonate and polyester track-etched membranes were tested by an independent laboratory with MEM Extract/L929 Mouse Fibroblast Cells. No evidence of cytotoxic response was noted during a seventy-two (72) hour exposure period, and the materials were judged to be negatively cytotoxic.

Q. Is there any way I can get rid of the pores on Sterlitech's Track-Etch filter membranes?

A. If you want to completely get rid of the pores, the easiest way to do so is to dissolve the filter away using chloroform or touch prep.

Q. How do I remove the PVP hydrophilic coating from the surfaces of a polycarbonate membrane filter?

A.  Sterlitech’s polycarbonate membranes are naturally hydrophobic but are coated with PVP to produce a hydrophilic version. While the full range of pores sizes (0.01um-30.0um) are available as hydrophilic filters, we are only able to offer a limited number of pore sizes (0.1um-10.0um) as hydrophobic (PVP-Free).

Fortunately, the PVP can be removed without much hassle.  Boil the membrane(s) for at least 1 hour in deionized ultra-pure water, remove, then air dry the membrane(s).  Once the membrane dries it will be hydrophobic and PVP-Free.

Q. Why are black polycarbonate membranes better than other membranes for counting bacteria?

A. Black polycarbonate membranes are better than cellulose membranes for the direct counting of bacteria, because they have a uniform pore size and a flat surface that retains all of the bacteria on top of the filter. Although cellulose filters also retain all of the bacteria, many become trapped inside the filter where they cannot be counted.

Thickness, Bubble Point, Pore Size, and Pore Density remain unchanged when stored at a temperature from +5 to 35°C and a humidity of 20 to 80 % for up to 8 years

Q. What is the thickness of the polypropylene membranes?

A. The thicknesses of our polypropylene membranes are:

  • 0.1um polypropylene: 75-110um
  • 0.2um polypropylene: 140-180um
  • 0.45um polypropylene: 140-180um

Q. What is the shelf life of your membranes?

A. Cellulose acetate (CA) and nitrocellulose (MCE) membranes will last 2 1/2 to 3 years before reverting to their natural hydrophobic states.  Nylon, on the other hand, is naturally hydrophilic so it does not have a shelf life.

An easy test for older membranes is to perform a "wetting out" test.  Dipping a small portion of the membrane into water (hydrophilic membranes) or an alcohol (hydrophobic membranes) works well.  If the membrane absorbs the material, or "wets out", it has not gone past the shelf life.

Yes, the cellulose acetate (CA) membranes contain a proprietary surfactant.  The surfactant is used to ensure the CA membrane filters have good wettability.

Yes, the polyether ether ketone (PEEK) membrane filters may be purchased with diameters other than 25 and 47mm.  Please contact [email protected] to inquire about availability and pricing.

Q. Which membranes are recommended for gravimetric analysis?

A. We have several membranes to recommend for gravimetric analysis.

  • Mixed Cellulose Esters (MCE) Membrane Filters, Plain: In gravimetric analysis using ashing techniques, (MCE) Nitrocellulose filters yield a residue of less than 0.045% of their initial weight.  They are hydrophilic with a non-cytotoxic wetting agent extractable level of less than 4% of their weight.
  • Polycarbonate Track-Etch Membranes (PCTE) - 25mm:  Polycarbonate Track-Etch or our Polyester Track-Etch (PETE) membranes are two membranes that offer exceptionally low tare weights, are non-hygroscopic, and exhibit extremely low absorption and adsorption losses.

Since these membranes are non-hygroscopic, they are particularly well suited for gravimetric analysis.  They do not require drying when used directly out of the package.  If they are wet, they can be dried rapidly and will not pick up moisture from the air during weighing.

    • Analytical Filter Funnels Funnels are available complete with low hold-up polypropylene support pads, gravimetric analysis is one of the ideal applications for this product.  Each package of 12 units comes with an extender that adapts the unit for side-arm flask filtrations.  Optional No. 8 rubber stoppers may be ordered for filtering with flask.


Q. What is the maximum operating temperature for the 17mm and 30mm syringe filters?

A.  The 17mm and 30mm syringe filters have a polypropylene housing, which allows them to withstand higher temperatures than an acrylic housing.  The 17mm and 30mm's have a maximum operating temperature of 180°C.  They are also able to be autoclaved.

No. It is a common misconception that sterile air/gas filtration requires the use of a filter containing small pores. A 0.2 µm filter is required for sterile liquid filtration; however, this pore size is not necessary when sterilizing air or gas since the filtration mechanisms in air differ significantly from those in liquid. For example, a filter that retains 0.3 µm aerosolized particles at 99.97% filtration efficiency can provide a sterile air/gas stream due to the mechanics of particle capture in the gas stream environment.

Typically, the criteria set by the ASTM (American Society for Testing of Materials) can be met using filtration media with pore sizes up to 1.0 µm; there are also published reports of filters that are an order of magnitude higher than captured particulate. The efficiency of a filter for the production of sterile (ultrapure) air is based upon the concentration of 0.3 µm Dioctylephthalate aerosol particles (DOP) that penetrate a filter at a predetermined flow and expressed as a percentage. A 99.97% DOP retention indicates a high efficiency particulate air (HEPA) grade rating (ASTM: D2986-71).

Ask An Expert

As an industry leader focused in unique micro and sub-micron filtration products, our goal is to support our customers by keeping them at the forefront of their industries. We're here to help with any filtration questions you might have so you can transform your ideas into reality, and tackle those big science problems. Feel free to reach out using the form below, our experts are ready to serve.