Sterlitech Sepa
CF II Membrane Cell System Specifications
Operating Parameters
Effective Membrane Area: 24 inch^2 (155 cm^2)
Hold-Up Volume: 2.4 oz (70 mL)
Maximum Pressure:
316SS Cell Body Only 1,000 psig (69 Bar);
Polypropylene and Acrylic Cell Bodies 100
psig (6.9 Bar)
Maximum Temperature:
316SS Cell Body 350°F (177°C);
Polypropylene Cell Body 180°F (82°C);
Clear Acrylic Cell Body 150°F (66°C)
pH Range: Membrane Dependent
Crossflow Velocity: Variable (See graph)
Feed Velocity vs. Flow
Rate

Superficial velocity = Q/(W*T)
Q is the flow rate
W is the width of the flow channel
T is the feed channel height
Connections
Feed, Concentrate: 1/4 inch FNPT
Permeate: 1/8 inch FNPT
Hydraulic Port for Piston: 1/4 inch FNPT
Pumps
Most lab scale pumps including piston-driven, diaphragm,
peristaltic, centrifugal and gear, are acceptable. Contact Sterlitech
for further recommendations.
Fluid Transfer
Depending on pressure, standard
tubing or hose can be used for connecting cells to pumps and vessels.
High-pressure hose should be used where the pressure is elevated.
Contact Sterlitech if additional help is required.
Material of Construction
Cell Body (316SS): 69 bar (1000 psig) max. operation
Holder: Anodized Aluminum
Concentrate Flow Control Valve: 316 Stainless Steel
O-Rings: Viton*
Concentrate Pressure Gauge: 316 Stainless Steel
Connections
Cell Body: 16.51 x 21.3 x 5.0 cm (6.50 x 8.38 x 2.04 in)
Holder: 20.0 x 28.0 x 20.0 cm (7.9 x 11.0 x7.9 in)
Shim: 14.6 x 9.5 cm (5.74 x3.74 in)
Feed Spacer: 14.6 x 9.5 cm (5.74 x3.74 in)
Membrane: 19.1 x 14.0 cm (7.5 x 5.5 in)
Permeate Carrier: 14.6 x 9.5 cm (5.74 x3.74 in)
Membrane Packs
Sterlitech manufactures a complete line of Sepa membrane
types for reverse osmosis, nanofiltration, ultrafiltration, and
microfiltration separations. For the Sepa-CF II system, Sterlitech
offers precut, individually wrapped membrane inserts in single membrane
type or membrane assortment packs.
Sepa CF Ordering
info.
Sepa CF Sample applications.
Sepa CF Principals of operation.
Sepa CF Benefits and Features.
Sepa CF FAQ's.
For Polymeric Membrane
info, please click here.
|